Alaris 8100 Infusion Pump from $17.58/mo

Alaris/CareFusion 8100 Infusion Pump Module

Alaris/CareFusion 8100 Infusion Pump Module
Product Specifications
BRAND  Alaris
MODEL  8100
BRAND Alaris
BRAND Alaris
TYPE Large Volume
EQUIPMENT CONDITION Refurbished
EQUIPMENT CONDITION Refurbished
DESCRIPTION
The BD CareFusion Alaris 8100 Pump Module is designed for the delivery of large volume fluids, medications, blood, and blood products using continuous intermittent delivery for adult, pediatric, or neonatal patients. Infusion delivery can occur via clinically acceptable routes of administration (e.g., intravenous, intra-arterial, subcutaneous, epidural, enteral, or irrigation of fluid spaces).

Clinicians can attach up to four pump modules to a single Point of Care (PC) unit allowing four different infusions. The CareFusion pump modules can be attached to either side of the PC unit to create the desired configuration.

The CareFusion Alaris Cardinal 8100 Pump module supports the Guardrails Suite MX software. This software helps to reduce the risk of medication errors by providing a test of reasonableness before the initiation of therapy.
ADDITIONAL INFORMATION
FEATURES:
  • LCD which displays the drug or IV fluid name, dose, and rate.
  • Indicators that display channel status-red for alarm, yellow for pause, and green for infusing.
  • Set-based, free-flow protection featuring the SmartSite needle-free valves.
  • Sets can be used for gravity infusion.
  • Designed for bag, bottle, or syringe delivery.
  • Supports primary and secondary “piggyback” delivery
SPECIFICATIONS:
  • Dimensions
    • Height: 8.9” (22.6 cm)
    • Width: 3.3” (8.4 cm)
    • Depth: 5.5” (14 cm)
    • Weight: 2.5 lbs (1.13 kg)
  • Accumulated Air Window
    • Single Bolus Setting: 50; Volume Window (mL): 2.8; Air That Causes Alarm: 10%
    • Single Bolus Setting: 75; Volume Window (mL): 8.0; Air That Causes Alarm: 20%
    • Single Bolus Setting: 250; Volume Window (mL): 8.0; Air That Causes Alarm: 30%
    • Single Bolus Setting: 500; Volume Window (mL): 12.0; Air That Causes Alarm: 30%
  • Bolus Volume, Maximum after Occlusion
    • Pressure Limit (mmHg): 50; Rate (mL/h): 25; Bolus Volume (mL): ≤0.3
    • Pressure Limit (mmHg): 525; Rate (mL/h): 25; Bolus Volume (mL): ≤0.6
  • Critical Volume
    • The maximum over-infusion which can occur in the event of a single fault condition is 0.6 mL.
  • Fluid Ingress Protection
    • IPX1, Drip Proof
  • Maximum Infusion Pressure
    • 654 mmHg (Maximum Occlusion Alarm Threshold plus tolerance)
  • Rate Accuracy
    • ±5% at rates between 1 and 999 mL/h; ±5.5% at rates <1 mL/h, 95% of the time with 95% confidence, under the conditions listed below.
    • Infusion Rate Range: 0.1 - 999 mL/h
    • Ambient Temperature: 68 ±4° F (20 ±2° C)
    • Source Container Height: 20 inches above top of Pump Module
    • Test Solution: Distilled Water
    • Distal Back pressure: 0 mmHg (0 kPa)
    • Needle: 18 gauge
    • Administration Set Model: 2210
  • Infusion of Air, Means to Protect Patient from
    • Ultrasonic Air-in-Line Detection: Maximum single bolus size = selectable 50, 75 or 250 microliters nominal (500 microliters in Anesthesia Mode)
  • IKVO (Keep Vein Open) Rate
    • Factory Default Setting is 1 mL/h if set rate is 1 mL/h or above; or set rate, if rate is 0.9 mL/h or below.
  • KVO Selection Range:
    • KVO rate can be set in System Configuration from 0.1 - 20 mL/h in 0.1 mL/h increments.
  • Occlusion Alarm Thresholds:
    • Pump Mode: 525 mmHg at rates ≥30 mL/h Varying level based on rate and patient back–pressure at rates <30 mL/h.
    • Selectable Mode: User selected, 50 - 525 mmHg in 25 mmHg increments.
  • Operating Principle:
    • Positive displacement
  • Shock Protection:
    • Type CF, Defibrillator Proof
  • Environmental Operating
    • Temperature Range: 41 to 104°F (5 to 40°C)
    • Relative Humidity: 20 to 90%
    • Atmospheric Pressure: 525 to 4560 mmHg (700 to 6080 hPa)
  • Storage / Transport
    • Temperature Range: -4 to 140°F (-20 to 60°C)
    • Relative Humidity: 5 to 85%
    • Atmospheric Pressure: 375 to 760 mmHg (500 to 1013 hPa)
RENT, FINANCE AND BUY OPTIONS

Rent for short-term needs, finance the purchase by leasing, or just purchase outright.

Offered by Jordan at Supplier #9626 in Tustin, CA
  Rental Rates   Finance Rates  Message Jordan

Don't see the specific model you need?

Request supplier quotes.

KWIPPED is an equipment rental and leasing marketplace. Suppliers have additional inventory available not listed on KWIPPED.
Our quoting and collaboration platform is an incredibly easy and efficient way to source from our entire supplier network.
Get quotes

Not sure if you should rent or finance?

Tell us what you're thinking and will offer a recommendation.

We have equipment rentals and leases for all your medical needs

First time using KWIPPED?

We've made it really easy.

Here are some resources to get you started with your first rental