Sample Compartment The sample compartment contains an aluminum block with 96 sample positions in an eight by twelve microplate format
A heated cover over the sample vessels eliminates the need for oil
Sample Tubes The sample compartment can hold:
96 MicroAmp Reaction Tubes (0.2-ml)
Electrical:
There are three versions of the GeneAmp PCR System 9600. The power requirements of these instrument versions are:
- 120 VAC, 60 Hz, 10 Amps, 15-amp slow blow fuse
- 220-240 VAC, 50 Hz, 5 Amps, 8amp slow blow fuse
- 100 VAC, 50-60 Hz, 12 Amps, 15-amp slow blow fuse
Control Panel The instrument control panel consists of the display, keypad, and indicator lights
Display The GeneAmp PCR System 9600 uses a 40-character blue luminescent display (20 characters x 2 lines).
The sample temperature is displayed in degrees Celsius to the nearest 0.1 degree C
Keypad The keypad consist of nine function keys and ten numeric keys
Indicator Lights: There are three indicator lights:
Heating
Cooling
Hot
Sample Temperature Range The temperature range of the sample block is: 4.0 degrees to 99.9 degrees C
Sample Temperature Accuracy The displayed sample temperature matches the average true sample temperature within +/-0.75 degrees C.
Displayed Sample Temperature The displayed sample temperature will match the specified setpoint temperature to within +/-0.2 degrees C, centered around the setpoint temperature two minutes after the sample temperature is within 1 degree C of setpoint
Sample Block Temperature Uniformity The sample block temperature uniformity is: +/-0.5 degrees C 30 seconds after the clock starts, over the range 50.0 to 99.9 degrees C
Sample Temperature Overshoot An average of 96 sample temperatures shows less than 0.5 degrees C overshoot for upramps and downramps
Ramp Time Repeatability Thermal setpoints are achieved within +/-5 seconds (not including the first cycle)
Sample Heating and Cooling Rate The sample temperature ramp rate for heating and cooling is approximately on degree C per second
The GeneAamp PCR System 9600 uses an RS-232C serial port as its printer interface. In addition, the following specifications apply: Baud Rate: 9600
Setup: no parity, 8 data bits, on stop bit
Dimensions:
Physical Dimensions of the instrument are: Height: 30.5 cm (12 in.)
Width: 52 cm (20 in.)
Depth: 61 cm (24 in.)
Weight: 40.5 kg (90 lbs.)